Efficacy of 99mTc-DTPA SPECT/CT in diagnosing Orbitopathy in graves’ disease

Abstract

Background

The most frequently used methods of assessing Graves’ orbithopathy (GO) include: Clinical Activity Score (CAS), ultrasonography (USG), computed tomography (CT), and magnetic resonance imaging (MRI). There exists another, slightly forgotten, imaging method: single-photon emission computed tomography (SPECT) with the use of diethylenetriaminepentaacetic acid tagged with 99mTc (99mTc-DTPA). These days it is possible to conduct a SPECT examination fused with a CT scan (SPECT/CT), which increases the diagnostic value of the investigation. The aim of this paper is to evaluate the usefulness of 99mTc-DTPA SPECT/CT in diagnosing Graves orbitopathy, as compared with other methods.

Methods

Twenty-three patients with suspected active (infiltrative-edematous) Graves’ orbithopathy were included in the study. Each patient underwent a CAS, an MRI, and a SPECT/CT. The obtained results were analysed statistically, with the assumed statistical significance of p < 0.05.

Results

The SPECT/CT and MRI were found to have the highest sensitivity: 0.93 each. The SPECT/CT had the highest specificity: 0.89. MRI and CAS had lower values: 0.78 and 0.56, respectively. The occurrence of an active form of GO had no impact on the exacerbation of exophthalmos or the thickness of the oculomotor muscles.

Conclusions

The 99mTc-DTPA SPECT/CT method provides a very good tool for assessing the active form of GO and can, alongside the MRI scan, be used as a referential diagnostic procedure in GO.

Peer Review reports

Background

Graves’ orbitopathy (GO) is a complex of ocular symptoms caused by autoimmune inflammation of the soft tissue in the orbital cavity. Active inflammation of the orbital cavity’s soft tissue initially presents only with swelling, but in the long term can lead to irreversible lesions in the form of fibrosis and steatosis [1].

Because the condition can resolve spontaneously, the decision to administer treatment should be preceded by an assessment of the intensity of inflammation. The most frequently used methods of evaluating Graves’ orbithopathy include: Clinical Activity Score (CAS), ultrasonography (USG), computed tomography (CT), and magnetic resonance imaging (MRI). There exists another, slightly forgotten, imaging method: single-photon emission computed tomography (SPECT) with the use of diethylenetriaminepentaacetic acid tagged with 99m Tc (99mTc-DTPA). The investigation involves intravenous (iv) administration of the 99mTc-DTPA radiotracer (used mainly in testing renal clearance), which then becomes extravasated at the site of inflammation, to combine with the polypeptides of the extracellular inflammatory fluid. The volume of 99mTc-DTPA’s accumulation at the site of inflammation, in this case in the soft tissue of the orbital cavity (mainly in the oculomotor muscles) is directly proportionate to the activity of the infiltrative and edematous process [2]. The method employs a gamma camera to capture gamma radiation produced by the 99mTc attached to DTPA, using the single-photon emission computed tomography (SPECT) technique. Modern gamma cameras are hybrid devices, combined with CT, which make it possible to conduct scintigraphy together with computed tomography (SPECT/CT), thus increasing the diagnostic accuracy of the test.

In view of the above, the purpose of this paper is to assess the efficacy of SPECT/CT scintigraphy using 99mTc-DTPA in diagnosing Graves’ orbitopathy, in comparison with other methods.

Methods

Patients

Twenty-three patients (women aged 25–60, 45 on average) with suspected active (infiltrative-edematous) Graves’ exophthalmos were included in the study. All the patients had been diagnosed with Graves’ disease, confirmed by the presence of above normal (norm: < 2 IU/l) levels of TRAb receptor antibodies in the blood serum. Fifteen of them were on antithyroid drugs in order to normalise serum levels of the thyroid hormones. The others were hypothyroid, post-radioiodine treatment and on thyroxine substitution therapy. The evaluation of GO involve using: the CAS score, the MRI test, and the SPECT/CT imaging technique. Each patient had these tests performed within 10 days after the commencement of the diagnostic process.

CAS score

The seven-point scale used for evaluating the clinical activity of GO is recommended by the European Group on Graves’ Orbitopathy (EUGOGO). It takes into account the following clinical symptoms, each of which represents one point on the scale: retroorbital pain, pain with eye movement, swelling of the eyelids, redness of the eyelids, conjunctival redness, conjunctival swelling, inflammation of the caruncle or plica. The presence of at least 3 out of the 7 above-mentioned clinical symptoms indicates an active form of GO [3].

MRI

The MRI of the orbital cavities was performed by means of a 3 T MRI scanner (Biograph mMR, Siemens, Germany), using the T1 sequence with and without intravenous contrast (Gadolin), as well as the T2 sequence. All the scans were conducted in the coronal and axial planes, with a slice thickness of 3 mm. The presence of infiltration and oedema within oculomotor muscles (T2 signal increased and T1 signal strengthened in inflamed muscles on contrast-enhanced scan) was taken as a positive result, whereas the absence of the active form of GO was considered to be a negative result [4].

SPECT/CT

The SPECT/CT of orbital cavities was conducted with the use of the radiotracer 99mTc-DTPA (PolAtom, Polska), given intravenously at the dose of 7 MBq/kg. The 3D tomography using a SPECT/CT gamma camera (Symbia T, Siemens, Germany) was performed at 20 min after radiotracer administration. The SPECT image was obtained through the application of iterative reconstruction. The voltage of the RTG lamp used for the scans equalled 80 kV and the anodic current - 30 mA. Then, the fused SPECT and CT (SPECT/CT) images with a slice thickness of 1.25 mm were analysed medically. Increased accumulation of the radiotracer in the muscles or in the oculomotor muscles, as compared with the background (i.e. the uptake of the radiotracer in the remaining craniofacial muscles) were taken as a positive result, i.e. presence of infiltrative-edematous lesions (Fig. 1).

Fig. 1
figure 1

99mTc-DTPA SPET/CT scintigraphy of the orbital cavity in coronal planes. All the rectus muscles of the eye are thickened: a) with increased accumulation of 99mTc-DTPA only in the medial and superior muscles of the right eye, which is indicative of active GO in the medial and superior muscles. b without increased accumulation of 99mTc-DTPA, which is indicative of non-active GO. The physiological 99mTc-DTPA uptake of nasal mucosa is visible at both cases. Abbreviations: MRI = magnetic resonance imaging; SPECT-CT = fused single photon emission computed tomography with computed tomography; GO = Graves’ orbitopathy; 99mTc-DTPA = diethylenetriaminepentaacetic acid tagged with 99mTc

As part of the SPECT/CT examination, an additional assessment was made of the thickness of the oculomotor muscles and the degree of the exophthalmos. The evaluation of the exophthalmos was performed in the standard manner on the basis of CT scans in the axial plane, obtained as a result of the SPECT/CT. The straight line joining the anterior tips of the zygomatic bones was the line of reference. The distance of > 23 mm from the line of reference to the front border of the eye was interpreted as indicative of exophthalmos. The thickness of the oculomotor muscles was measured using the scans obtained as a result of the SPECT/CT examination [5].

Statistical analysis

The statistical analysis of the study results was performed using Statistica 13.1 software (Stat Soft, Tulsa, USA).

To assess the usefulness of the diagnostic methods (CAS, MRI, SPECT/CT) in detecting the active form of GO, contingency tables were created, which included such measures as: sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and accuracy (ACC).

The coefficient of confidence for each measure was calculated on the basis of the Clopper-Pearson method for an individual proportion and equalled 0.95.

The evaluation of the dependence between the degree of exophtalmos and the overall thickness of the oculomotor muscles and the presence of an active inflammatory form of GO, as well as the comparison of the studied methods (CAS, MRI, SPECT/CT) was conducted by means of one-way analysis of variance.

Results

When analysing the results of CAS, MRI, and SPECT-CT in detecting an active form of GO in the examined patients, one should bear in mind that the diagnoses were not always consistent (Table 1). Therefore, the following assumptions were adopted in order to perform the statistical analysis:

  • result of the test was taken as truly positive (active form of GO) if at least two of the compared methods yielded positive results;

  • result of the test was taken as truly negative (non-active form of GO) if at least two of the compared methods yielded negative results;

  • result of the test was taken as falsely negative if only one of the compared methods yielded a negative result;

  • result of the test was taken as falsely positive if only one of the compared methods yielded a positive result;

Table 1 Results of active (+) and non-active (−) forms of GO in 23 patients after CAS, MRI, and SPECT/CT

Next, taking into consideration the above assumptions, contingency tables were constructed for the particular diagnostic methods: CAS, MRI, and SPECT/CT (Tables 2, 3, and 4).

Table 2 Correlation between real occurrence of active or non-active GO and result of CAS (contingency table)
Table 3 Correlation between real occurrence of active or non-active GO and result of MRI (contingency table)
Table 4 Correlation between real occurrence of active or non-active GO and result of SPECT/CT (contingency table)

It was demonstrated that of the three analysed diagnostic methods, SPECT/CT and MRI had the highest sensitivity: 0.93. Therefore, 93% of the patients with the active form of GO were positively diagnosed by means of these methods. CAS proved to have a lower sensitivity, namely 0.86.

The SPECT/CT scan was found to have the highest specificity, because as many as 89% of the patients without an active form of GO were correctly diagnosed using this method, i.e. had negative results, whereas in the case of MRI the figure proved to be 78%, while in the case of CAS - only 56%. A patient with a positive result of a SPECT/CT scan can have a 93% certainty that he or she has an active form of GO. These values are significantly lower for MRI and CAS: 87 and 75%, respectively. The SPECT/CT method seems superior also in the case of a negative result. A patient can then be 89% certain that he or she does not have an active form of GO, whereas negative results of MRI and CAS mean a certainty of 88 and 71%, respectively. It should also be added that SPECT/CT ensures the highest accuracy (i.e. probability of a correct diagnosis). A patient who undergoes a SPECT/CT scan (regardless of the obtained result) can be 91% certain of the accuracy of the diagnosis, while the same figure equals 87 and 74% in the cases of MRI and CAS, respectively (Table 5).

Table 5 Assessment parameters for diagnostic methods CAS, MRI and SPECT/CT in detecting active GO

The occurrence of the active form of GO did not have an impact on the deterioration of exophthalmos or the thickness of the oculomotor muscles. The total thickness of the oculomotor muscles in one eye in the active and non-active forms of GO equalled, respectively, 40 mm +/− 6 mm and 39 mm +/− 7 mm, while the degree of exophthalmos was, respectively, 23 mm+/− 5 mm and 23 mm +/− 4.5 mm. The measurement differences were not statistically significant, p = 0.09 and p = 0.12 (Table 6).

Table 6 Impact of active GO on average total thickness of muscles in orbital cavity and occurrence of exophthalmos

Discussion

The role of 99mTc-DTPA scintigraphy of the orbital cavity in diagnosing the active (infiltrative-edematous) form of GO has been known for many years. The first mention of the method’s application appeared in 2002. Galuska et al. proved that 99mTc-DTPA SPECT can be used to evaluate the autoimmune inflammation of the retrobulbar area in GO patients. Increased accumulation of the radiotracer in the orbital cavity, as compared to the background, indicated the presence of local inflammation. They made it clear, however, that because of the small number of examined patients (only 9), a precise evaluation of the method’s usefulness was impossible [6]. Authors from the same research institution, having conducted a study on a larger number of patients (21 persons), concluded that SPECT scintigraphy using 99mTc-DTPA could be a supplementary diagnostic approach helpful in deciding whether immunosuppressive treatment should be commenced and facilitating the assessment of this treatment’s efficacy for treating GO. Like most authors, they regarded MRI as the first step in the diagnostic algorithm of GO [7]. Also Szabados writes about the usefulness of 99mTc-DTPA SPECT scintigraphy in GO therapy. Authors have proved that in patients with a lack or low levels of 99mTc-DTPA accumulation in the orbital cavity (in other words if there is no autoimmune inflammation or if it is mild), treatment with radiation did not bring any beneficial therapeutic effect. Therefore, a scintigraphic examination can facilitate the decision as to whether GO should be treated with systemic medications (steroids) or local methods (radiotherapy, surgery). It can also be helpful in monitoring the efficacy of treatment [2, 8].

The authors of the above studies drew their conclusions based on scintigraphic examination of orbital cavities performed using the SPECT method. This tomographic technique makes it possible to create a three-dimensional image with the use of gamma radiation emitted by a radiotracer accumulated in the organism (in this case, the orbital cavity). One advantage of this method is that it produces a spatial visualisation of the biological activity of the examined area. Speaking of biological activity in a SPECT scan of orbital cavities, we mean the detection of the presence or absence of the active form of GO (details in the introduction to this paper). A disadvantage of this technique is the relatively low resolution of the scan (> 1 cm), in comparison with other imaging methods, such as CT or MRI (where resolution < 1 mm), which makes interpretation of the result more difficult [9,10,11]. SPECT imaging only enables one to state whether an active form of GO is present or absent in the retro-orbital area, without allowing for a more precise indication of which of the oculomotor muscles is affected by inflammation. Besides, the presence of physiologically increased accumulation of the radiotracer in the mucous membranes of the adjacent lateral nasal sinuses can sometimes be mistakenly interpreted as increased accumulation of the radiotracer in the retro-orbital cavity, thus leading to a falsely positive result. However, our studies, aimed at evaluating the form of GO, use a hybrid method: SPECT/CT. In comparison with the traditional SPECT, the SPECT/CT technique provides a much more detailed image of the location of the accumulated radiotracer, thus improving the precision of the interpretation of the image as either physiological or pathological, depending on its exact position. This is confirmed by the very high sensitivity and specificity of the SPECT/CT method, demonstrated in our paper and equalling 93 and 89%, respectively. These values are comparable with those of MRI as far as sensitivity is concerned, and are higher as regards specificity. In our study, the sensitivity of the MRI tests was found to be 93%, whereas specificity equalled 78%. Such a high specificity of scintigraphic scans can be explained by the fact that it is capable of detecting abnormal quantities of substances at the level of picomoles, while other imaging methods (CT, MRI) can only detect millimoles [12]. In view of the above, in our study, a positive result of a SPECT/CT scan of one patient, given negative results of CAS and MRI, does not necessarily have to mean a falsely negative result, as was assumed for the sake of the statistical analysis (Tables 1 and 2). The patient can simply have the active form of GO of low severity, impossible to detect with CAS and MRI. On the other hand, a single falsely negative result can result from the difficulty of assessing the accumulation of the radiotracer in the medial oculomotor muscles due to the presence of physiologically increased build-up of the radiotracer in the adjacent mucosa of the nasal sinuses. However, the number of patients we have examined is too low to allow us to state this with absolute certainty.

What is also interesting is that our study suggests a lack of statistically significant difference between the active and non-active forms of GO taking into consideration the severity of exophthalmos and the thickness of the oculomotor muscles. This can be explained by the relatively small group of investigated patients, but also by the fact that no norms exist for the width of muscles in the active or non-active forms of GO. Frequently, the muscles of a patient with active GO are smaller in volume than those of a patient with non-active GO, and vice versa. It is similar with the severity of exophthalmos. In our study, this is reflected in the low specificity of CAS in diagnosing the active form of GO, equalling just 56%. The reason is that in order to differentiate the forms of GO, the method bases on the evaluation of clinical symptoms manifesting mainly as an expansion of the oculomotor muscles and the occurrence of exophthalmos.

Finally, it should be emphasised that the SPECT/CT examination using 99mTc-DTPA is safe in terms of radiation protection. The overall dose of ionising radiation absorbed by the patient is the sum of: 1mSV received during the SPECT scan and about 0.6 mSV received during the low-dose CT. It is, therefore, only a small fraction of the absorbed dose of ionising radiation which is universally regarded as safe for healthy persons who work with ionising radiation, i.e. 20mSV per year [13, 14]. Moreover, taking into account the economic aspects, it should be noted that the method is less expensive than MRI, the standard investigation in this condition.

Conclusion

On the basis of our analysis, it can be concluded that the 99mTc-DTPA SPECT/CT method provides a very useful tool for evaluating the active form of GO, in many respects superior to the routinely applied methods. It seems that it can be used as a referential diagnostic technique in GO, along with MRI.

Abbreviations

99mTc-DTPA:

Diethylenetriaminepentaacetic acid tagged with 99mTc

ACC:

Accuracy

CAS:

Clinical Activity Score

CT:

Computed tomography

GO:

Graves’ orbithopathy

MRI:

Magnetic resonance imaging

NPV:

Negative predictive value

PPV:

Positive predictive value

SPECT:

Single-photon emission computed tomography

SPECT/CT:

SPECT examination fused with a CT scan

TRAb:

TSH receptor antibody

TSH:

Thyroid-stimulating hormone

USG:

Ultrasonography

References

  1. Bahn RS. Graves’ ophthalmopathy. N Engl J Med. 2010;362(8):726–38.

    CAS  PubMed  PubMed Central  Google Scholar 

  2. Galuska L, Barna SK, Varga J, Garai I, Nagy EV. The role of 99mTc-DTPA retrobulbar SPECT in staging and follow-up of Graves’ orbitopathy. Nucl Med Rev Cent East Eur. 2018;21(1):54–8.

    Article  PubMed  Google Scholar 

  3. Bartalena L, Baldeschi L, Dickinson A, Eckstein A, Kendall-Taylor P, Marcocci C, et al. Consensus statement of the European group on Graves’ orbitopathy (EUGOGO) on management of GO. Eur J Endocrinol. 2008;158(3):273–85.

    Article  CAS  PubMed  Google Scholar 

  4. Yokoyama N, Nagataki S, Uetani M, Ashizawa K, Eguchi K. Role of magnetic resonance imaging in the assessment of disease activity in thyroid-associated ophthalmopathy. Thyroid. 2002;12(3):223–7.

    Article  PubMed  Google Scholar 

  5. Kirsch E, von Arx G, Hammer B. Imaging in Graves’ orbitopathy. Orbit. 2009;28(4):219–25.

    PubMed  Google Scholar 

  6. Galuska L, Leovey A, Szucs-Farkas Z, Garai I, Szabo J, Varga J, et al. SPECT using 99mTc-DTPA for the assessment of disease activity in Graves’ ophthalmopathy: a comparison with the results from MRI. Nucl Med Commun. 2002;23(12):1211–6.

    Article  CAS  PubMed  Google Scholar 

  7. Galuska L, Leovey A, Szucs-Farkas Z, Szabados L, Garai I, Berta A, et al. Imaging of disease activity in Graves’ orbitopathy with different methods: comparison of (99m)Tc-DTPA and (99m)Tc-depreotide single photon emission tomography, magnetic resonance imaging and clinical activity scores. Nucl Med Commun. 2005;26(5):407–14.

    Article  PubMed  Google Scholar 

  8. Szabados L, Nagy EV, Ujhelyi B, Urbancsek H, Varga J, Nagy E, et al. The impact of 99mTc-DTPA orbital SPECT in patient selection for external radiation therapy in Graves’ ophthalmopathy. Nucl Med Commun. 2013;34(2):108–12.

    Article  PubMed  Google Scholar 

  9. Khalil MM, Tremoleda JL, Bayomy TB, Gsell W. Molecular SPECT imaging: an overview. Int J Mol Imaging. 2011;2011:796025.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Huda W, Abrahams RB. X-ray-based medical imaging and resolution. AJR Am J Roentgenol. 2015;204(4):W393–7.

    Article  PubMed  Google Scholar 

  11. Pooley RA. AAPM/RSNA physics tutorial for residents: fundamental physics of MR imaging. Radiographics. 2005;25(4):1087–99.

    Article  PubMed  Google Scholar 

  12. Gotthardt M, Bleeker-Rovers CP, Boerman OC, Oyen WJ. Imaging of inflammation by PET, conventional scintigraphy, and other imaging techniques. J Nucl Med Technol. 2013;41(3):157–69.

    Article  PubMed  Google Scholar 

  13. Montes C, Tamayo P, Hernandez J, Gomez-Caminero F, García S, Martín C, et al. Estimation of the total effective dose from low-dose CT scans and radiopharmaceutical administrations delivered to patients undergoing SPECT/CT explorations. Ann Nucl Med. 2013;27(7):610–7.

    Article  CAS  PubMed  Google Scholar 

  14. Delves D, International Atomic Energy Agency. IAEA safety glossary: terminology used in nuclear safety and radiation protection. 2007. Vienna: International Atomic Energy Agency; 2007. p. 227.

    Google Scholar 

Download references

Acknowledgments

We are grateful to all of the staff for their help with data collection and for technical assistance.

Funding

The authors declare they did not receive any funding for this study.

Author information

Authors and Affiliations

Authors

Contributions

PS contributed to conception and design, acquisition of data, analysis and interpretation of data, drafting of the manuscript, statistical analysis. SA, MM and MS performed the statistical analysis. LZ, KS and KM contributed to acquisition of data. AP and JM made an important intellectual contribution by taking part in drawing substantive conclusions in the paper. All authors have read and approved the final manuscript.

Corresponding author

Correspondence to Piotr Szumowski.

Ethics declarations

Ethics approval and consent to participate

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964. Helsinki declaration and its later amendments or comparable ethical standards.

The study was approved by the bioethical commission of Medical University of Bialystok, Poland. All participants provided written informed consent to participate.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Additional information

Paper financed from statutory activity funds of the Medical University of Bialystok

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Szumowski, P., Abdelrazek, S., Żukowski, Ł. et al. Efficacy of 99mTc-DTPA SPECT/CT in diagnosing Orbitopathy in graves’ disease. BMC Endocr Disord 19, 10 (2019). https://0-doi-org.brum.beds.ac.uk/10.1186/s12902-019-0340-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://0-doi-org.brum.beds.ac.uk/10.1186/s12902-019-0340-0

Keywords